UV1901 dual-beam UV-visible spectrophotometer, the dual-beam system has low noise, low stray light, and a bandwidth of 1nm with higher accuracy. Detachable sample chamber supports multi-wavelength testing, spectral scanning, quantitative and kinetic detection, USB output, large-screen display, and stable and high-precision operation at all wavelengths.

86-15688870591
cedric@cedriccn.com
Description

Product Features:

·Dual beam metering system, combined with advanced circuit measurement and control systems, makes the instrument highly stable and extremely low noise.

·The newly designed excellent optical system and high-performance holographic flash grating ensure the low stray light of the instrument, allowing the instrument to operate stably at the full wavelength in four decimal places after the absorbance value (2 decimal places after the transmittance value).

·The sample chamber design with a detachable structure makes it easy to replace different accessories to meet different analysis needs.

·Photometric measurement: The transmittance and absorbance at 1 to 6 wavelengths can be measured simultaneously.

·Spectroscopic measurement: Scan the transmittance, absorbance and energy in the wavelength range, and various data processing can be carried out such as peak and valley detection, derivative operation, graph operation, etc.

·Quantitative measurement: single-wavelength, double-wavelength, three-wavelength and multi-wavelength measurement of 1~9 working curve (1~4 times) regression.

·Kinetic measurement: Time scanning of transmittance and absorbance at any set wavelength and various data operations can be performed.

·Data output: Data files and parameter files can be accessed, and the measurement results can be output in a standard common data file format.


Product Specification:

Model

UV1901

Wavelength range

190-1100nm

Spectral bandwidth

1nm

Wavelength accuracy

±0.1nm (at 656.1nm), ±0.3nm in the whole area

Wavelength repeatability

≤0.1nm

Photometric accuracy

±0.002A (0-0.5A), ±0.004A (0.5-1A), ±0.2%T (0-100%)

Photometric repeatability

±0.001A (0-0.5A), ±0.002A (0.5-1A), ≤0.15%T (0-100%)

Stray light

≤0.02%T

Stability

±0.0004A (at 500nm)

Baseline straightness

±0.0004A

Working method

T, A, C, E

Photometric range

0-200%T,-4-4A

Data output

USB interface

Printout

Parallel port

Display system

320×240-bit highlighting, large-screen LCD

Packing Size

910*710*530mm

Weight

31kg/47kg