The L-512 tablet dissolved oxygen meter is equipped with a 5-inch LCD screen, the body is as thin as 15mm, and uses a 32-bit processor and capacitive touch. Supports automatic temperature/salt/air pressure compensation, fast response, good stability, and can measure DO and saturation. The 3.5mm electrode interface is easy to use, can store 300 sets of data, is USB rechargeable, has long lithium battery life, and has IP57 protection, making it suitable for field and laboratory use.
86-15688870591
cedric@cedriccn.com
Product Features:
·5-inch LCD screen, compact stand and rotating electrode holder. It has a small size with a thickness of only 15mm, suitable for on-site, field or laboratory use. Equipped with a built-in ARM 32-bit microprocessor chip, the circuit board adopts SMT (Surface Mount Technology), and the buttons use CTS (Capacitive Touch Sensing) technology.
·Different pH electrodes can be replaced easily and quickly (matching electrodes can be selected for measuring different samples) to measure pH.
·3.5mm electrode connecting wire for simple connection and easy use (electrode length can be customized according to customer needs).
·Automatic temperature compensation and automatic standard solution recognition. For pH measurement, it can automatically recognize 3 groups of standard solutions (15 types of buffer solutions) and support 1-3 point automatic calibration.
·It can store 300 data logs, including readings, GLP data, date and time in the data settings.
Product Specification:
Model | L-512 | |
DO | Measuring range | (0.00 - 20.00) mg/L(ppm) |
Resolution | 0.01/0.1 | |
Accuracy | Electrometer:±0.10 mg/L,Meter:±0.40 mg/L | |
Dissolved oxygen saturation | 0.0 - 200% | |
Saturation resolution | 0.1% | |
Saturation accuracy | ±2.0%FS | |
Temperature compensation range | (0 - 100)℃ (Automatic) | |
Salinity compensation range | (0 - 45) ppt | |
Air pressure Compensation range | (0 - 200) kPa | |
Air pressure resolution | 0.1 | |
Air pressure accuracy | ±2% | |
Temperature | Range | (0 - 100)℃ |
Resolution | 0.1℃ | |
Accuracy | 5-60℃:±0.4℃,Others:±0.8℃ | |
Other Parameters | Display | 5-inch LCD screen |
Data storage | 300 sets | |
Stability | (±0.07 mg/L)/1 h | |
Response time | ≤30 s (90% response at 25℃) | |
Power | 3.7 V/1800 mA rechargeable lithium battery | |
Communication interface | Standard USB interface | |
Dimensions & Weight | (190×130×15) mm/300 g | |
Working Conditions | Temperature | 5 - 40℃ |
Humidity | 5 - 85% | |
IP grade | IP57 | |
Packing Size | 500*450*120mm | |
Weight | 0.5kg/2kg | |
Fil the form and send us your message.Get in touch with PAULAIR and find a specialized consultant for you !