Tel:86-15688870591 E-mail:cedric@cedriccn.com
Dual-beam UV-visible spectrophotometer U9
Dual-beam UV-visible spectrophotometer U9

The U9 dual-beam UV-visible spectrophotometer uses a long-life pulse xenon lamp and requires no preheating. 10.1-inch large touch screen, dual-beam optical system, wavelength 190-1100nm, bandwidth 1/2nm optional. 32GB storage, supports USB/printing/multi-format export, stable and reliable, suitable for high-precision detection.

86-15688870591
cedric@cedriccn.com
Description

Product Features:

·Fast and convenient, the instrument is ready to use without preheating. The instrument uses pulsed xenon lamps as light source, and the pulsed xenon lamps with 1 billion flashes can last for 7 years.

·10.1-inch color touch display, the ultra-large screen display directly displays various scanning curves and maps, allowing users to complete all measurement needs without the help of a computer.

·Supports USB disk storage, which is convenient for users to use. The data measured by users can be exported directly to USB disk, which can support excel, txt format, and picture format (four formats can be output: *.csv, *.qua.*.tet, *.bmp).

·Data output: RS-232C serial port (print), USB drive (online), USB HOST (connected to USB disk), and comes with 32GB storage standard.

·The industry's first advanced 32-bit Cortex_M3 processor, with a main frequency of 120M, and the instrument can store 5,000 test data or 500 working curves.

·Suspension optical system design, strengthen the design of thickened aluminum base plate to eliminate the impact of vibration or deformation on the optical system; double-layer design, completely separate the circuit parts of the optical path.

·The instrument photoelectric signal detection device makes the instrument signal-to-noise ratio lower and the instrument is more stable.


Product Specification:

Model

U9

Data display

10.1 inch 1024×800 color capacitive touch screen

Optical system

Dual beam, Built-in reference detector

Light source

Flashing xenon lamp

Receiver

Silicon Photodiode

Spectral bandwidth

1nm or 2nm optional

Wavelength range

190-1100nm

Wavelength repeatability

0.1nm

Wavelength accuracy

±0.3nm

Wavelength resolution

0.1nm

Stray light

≤0.05%T at 220nm and 360nm

Photometric method

Transmittance/absorbance/concentration/energy

Analysis software

Support