UV1810 large-screen scanning UV spectrophotometer, equipped with ARM system, large-screen LCD display, supports USB storage and printing. The optical system is excellent, the suspended optical path design is stable, and it is equipped with an eight-link automatic sample rack, which can complete multiple types of detection, with high measurement accuracy and fast response.

86-15688870591
cedric@cedriccn.com
Description

Product Features:

·Adopt the latest ARM system, built-in Chinese and English operating systems, and can store and transfer data through the USB interface

·Powerful function, the host can independently complete the functions of photometric measurement, quantitative weighing, spectral scanning, kinetics, DNA/protein testing, multi-wavelength testing and data printing.

·Design a unique optical system, high-performance 1,200 bar/mm grating and imported receiver ensure that the instrument has excellent performance indicators.

·24-bit high-speed, high-precision A/D conversion, high instrument accuracy and fast response speed.

·A wide sample chamber that can accommodate cuvettes of 5-100mm of various specifications.

·You can directly connect the printer to print maps and experimental data.

·The instrument adopts suspension optical system design and long-range optical path design, which greatly improves the stability and reliability of the instrument and the optical resolution of the instrument.

·The instrument is equipped with a disc rotary automatic eight-connected sample holder, which has good reliability and centered light spot (compared to the double-row automatic eight-connected frame), so that the instrument has high metering accuracy.


Product Specification:

Model

UV1810

Wavelength range

190-1100nm

Spectral bandwidth

1nm

Wavelength repeatability

≤0.1nm

Photometric accuracy

±0.002A (0-0.5A), ±0.004A (0.5-1A), ±0.2%T (0-100%)

Photometric repeatability

±0.001A (0-0.5A), ±0.002A (0.5-1A), ≤0.15%T (0-100%)

Stray light

≤0.03%T

SStability

±0.0005A (at 500nm)

Working method

T, A, C, E

Photometric range

0-200%T,-4-4A

Data output

USB interface

Printout

Parallel port

Display system

320×240-bit highlighting, large-screen LCD

Light source

Tungsten lamps and deuterium lamps

Detector

Silicon photodiode

Noise level

0.05%T

Packing Size

820*670*390mm

Weight

21kg/26kg