1. Digital display measurement display value.2. Automatic zero adjustment and automatic adjustment of 100.3. The wide sample chamber can accommodate 5-100mm cuvettes.4. Optional output port RS232 can realize the function of connecting to the printer and the computer.5. Use original imported tungsten lamps to ensure the service life of the instrument.6. The instrument adopts the latest microcomputer processing technology and is easy to operate.7. Automatic optical doors ensure the service life of photoelectric sensors, making instrument testing easier.
Specification:
Model | C-722N |
Optical system | Achromatic aberration (1200 bars/mm grating) |
Wavelength range | 350-1020nm |
Wavelength accuracy | ±2nm |
Spectral bandwidth | 4nm |
Wavelength repeatability | ≤0.5nm |
Transmission ratio accuracy | ±0.5%T |
Stability | ±0.004A/h@500nm |
Drift | ≤0.1%T |
Noise | ≤0.2%T |
Working method | T, A, C |
Zero adjustment method | Automatic |
Weight | 8kg |
Output method | Printer |
Light source | Tungsten lamp |
Printing interface | Optional |
Photometric range | 0-200%T, -3-3A, 0-9999C |
Fil the form and send us your message.Get in touch with PAULAIR and find a specialized consultant for you !