MT series ultrapure water system, output 10-30L/h, maximum resistivity 18.2MΩ·cm. It adopts stainless steel precision filtration and nano-antibacterial water storage bags, and LCD monitors water quality online. Low TOC and endotoxin removal configurations are optional, and it is suitable for experimental scenarios such as HPLC and molecular biology.
86-15688870591
cedric@cedriccn.com
Application:
MT: physical&chemical test, blood inspection, flushing, biochemical, micro, HPLC, IC and GC analysis.
MT-V: atom absorption/emission mass spectrum analysis, TOC detection, environment monitoring, biochip developing, micro organics analysis
MT-F: ICP/MS, PCR, PPT. isotope analysis, molecular biology. life science,disease control center, medicine and quality inspect institute, college research lab etc.
MT-VF: molecular biology, life science,gene study, cell culture, amino acid analysis, protein purify, medical inspection.
Product Features:
·Built-in filter 10 micron 316L stainless steel precision mesh filter to remove fine impurities instead of PP cotton filter;manual drain valve, automatic cleaning at any time, soft scraping design for more thorough drainage, transparent shell filtering effect is intuitively visible; filter element is free to save money and effort
·It removes residual chlorine, heavy metals, and inhibits the growth of fungi at the front end of the purified water. It has a long service life,better protects the post-purification system, and extends the use time of the ultra-purification column.
·The water quality of the innovative built-in transparent pressure bucket is transparent,and the innovative replaceable liner water storage bag uses nano antibacterial technology to design the material to resist special ultraviolet rays reduce the odor of microorganisms and bacteria, and reduce organic chemicals such as nitrite. Prevent secondary water pollution in the pressure tank
·The case adopts plastic spraying process and side-opening door design. The filtering system adopts 11-inch U-shaped quick-inserted filter column for easier replacement and maintenance, and has the function of upgrading water quality and quantity.
·Backlight LCD display, dual waterway online monitoring. which can simultaneously display RO conductivity value, UP resistivity value and equipment operating status indication.
·Menu-type operation, which can modify and set a number of measurement parameters; has a number of operating status displays and indicator lights to promptly reflect the operating status of the equipment, high-pressure shutdown,
low-pressure alarm, water full shutdown, etc.
·RO water Ultra-pure water water quality alarm warning prompt Automatic replacement of the filter element, you can set the water with drawal time; touch the button to take water, RO UP water light indicator automatically prompts.
·Equipped with a portable test pen, which can test TDS content, conductance and water temperature anytime anywhere
·Pipeline adopts a quick-plug interface, the pipe joint passes NSF certification, dual-wavelength UV lamps, effectively sterilizes and reduces TOC, and the original imported 5000D ultrafiltration module effectively removes endotoxin.
MT SERIES Filtering Flow Chart
(The highest configuration MT-VF for reference)

Product Specification:
Model | Standard | Eliminating endotoxin | Low TOC | Synthesizing |
MT-10 | MT-F10 | MT-V10 | MT-VF10 | |
MT-15 | MT-F15 | MT-V15 | MT-VF15 | |
MT-20 | MT-F20 | MT-V20 | MT-VF20 | |
MT-30 | MT-F30 | MT-V30 | MT-VF30 | |
Process | PRE+KDF+PP+AC+ RO+UP+TF | PRE+KDF+PP+AC+ RO+UP+UF+TF | PRE+KDF+PP+AC+ RO+UV+UP+TF | PRE+KDF+PP+AC+ RO+UV+UP+UF+TF |
Output(25°℃)* | 10/15/20/30L/H | |||
Flow rate | Up to 2 liters/minute(with pressure tank) | |||
Pure water outlet | 2:reverse osmosis water,ultrapure water | |||
UItra pure water quality | ||||
Resistivity(25°℃) | 18.2MΩ.cm | |||
TOC* | <10ppb | <10ppb | <3ppb | <3ppb |
Bacteria | <0.1cfu/ml | |||
Particle(>0.22μm) | <1/ml | |||
Endotoxin | N/A | <0.001Eu/ml | N/A | <0.001Eu/ml |
RO water quality | ||||
lon rejection rate | 95%-99%(new RO membrane) | |||
Particles and bacteria rejection rate | >99% | |||
Feed water requirements | Tap water,temperature:5-45°℃,pressure:0.15-0.4Mpa | |||
Dimension and weight | Length×Width×Height:47×45×55cm/Weight:about 30KGS | |||
Electrical requirements | AC110-240V,50/60Hz | |||
Power | 48-240W | |||
Standard configuration | Machine(including 1 set of purification column)+terminal filter+accessory package | |||
Fil the form and send us your message.Get in touch with PAULAIR and find a specialized consultant for you !